User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Wavetek:Model 75 S-1190 Sub 1 DATE: 28-Apr-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 16 NUMBER OF LINES: 220 CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 HEAD {} 1.004 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.005 HEAD {º Amplitude Test º} 1.006 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.007 HEAD {}ÍÍ Amplitude Test ÍÍ 1.008 JMP 2.001 1.009 EVAL Dummy 2.001 DISP Set the UUT MENU as follows: 2.001 DISP SETUP .................................... press 3 times 2.001 DISP F2 ................................ (RSET) press 1 time 2.001 DISP F3 .................................(EXEC) press 1 time 2.002 DISP View the UUT display and verify display reads: 2.002 DISP RESET (Ver 2.2S)X 2.003 DISP Set the UUT MENU as follows: 2.003 DISP EDIT ..................................... press 2 times 2.003 DISP F1 ................................ (FUNC) press 1 time 2.004 DISP Rotate the UUT CONTROL KNOB clockwise until display 2.004 DISP reads "FUNC SINE". 2.005 DISP Set the UUT MENU as follows: 2.005 DISP F3 ................................. (EXEC) press 1 time 2.006 IEEE [@1281]ACV AUTO,ACCP 2.007 DISP ÿ Connect the UUT and the 1281 as follows: 2.007 DISP ÿ [27][91]1m MODEL 75 TO 1281 2.007 DISP ÿ NOTE! USE A 50 ê FEEDTHROUGH 2.007 DISP ÿ ÚÄÄÄÄÄ¿ ÚÄ INPUT HI 2.007 DISP ÿ ARB OUTPUT ÍÍÍÍÍͳ 50ê Æ͵ 2.007 DISP ÿ ÀÄÄÄÄÄÙ ÀÄ INPUT LO 2.008 IEEE [@1281]RDG?[I] 2.009 ACC 20 3.53V 130P% 10P/ 2.010 MATH MEM=(MEM*2.828) 2.011 MEMC 10.0V 1U 488.28H 3.001 IEEE [@1281]FREQ?[I] 3.002 ACC 20 488.28H 130P% 5P/ 3.003 MEME 3.004 MEMC H 10U 10Vp-p 4.001 HEAD Amplitude Test 4.002 DISP ÿ Connect the UUT and the 1281 as follows: 4.002 DISP ÿ [27][91]1m MODEL 75 TO 1281 4.002 DISP ÿ NOTE! REMOVE 50 ê FEEDTHROUGH 4.002 DISP ÿ ÚÄ INPUT HI 4.002 DISP ÿ ARB OUTPUT ÍÍÍÍÍÍÍÍÍÍÍÍÍ͵ 4.002 DISP ÿ ÀÄ INPUT LO 4.003 DISP Set the UUT MENU as follows: 4.003 DISP PARAMATER ................................ press 2 times 4.003 DISP F1 ..................................(AMPL) press 1 time 4.004 DISP Rotate the UUT CONTROL KNOB CCW until display 4.004 DISP reads "AMP 25.00 mVp". 4.005 DISP Set the UUT MENU as follows: 4.005 DISP EDIT ..................................... press 4 times 4.005 DISP F2 ................................. (STOP) press 1 time 4.005 DISP Rotate CONTROL KNOB for a SP=0000 display. 4.006 DISP Set the UUT MENU as follows: 4.006 DISP EDIT ..................................... press 2 times 4.006 DISP F2 .............................. (ADJX,Y) press 2 times 4.007 DISP Adjust CONTROL KNOB for a UUT display of Y= 2047. 4.008 IEEE [@1281] DCV AUTO,FAST_ON 4.009 IEEE [@1281] RDG?[D2000][I] 4.010 MATH M[1] = MEM 4.011 DISP Adjust CONTROL KNOB for a UUT display of Y=-2047. 4.012 IEEE [@1281] RDG?[I] 4.013 MATH M[2] = MEM * -1 4.014 MATH MEM = M[2] + M[1] 4.015 MEM* 1000 4.016 ACC 200 50.00mV 14P% 0.5P/ 4.017 MEME 4.018 MEMC mVp-p 0.95U 25mVP 5.001 DISP Set the UUT MENU as follows: 5.001 DISP PARAMATER ................................ press 2 times 5.001 DISP F1 ..................................(AMPL) press 1 time 5.002 DISP Rotate the UUT CONTROL KNOB clockwise until display 5.002 DISP reads "AMP 250.0 mVp". 5.003 DISP Set the UUT MENU as follows: 5.003 DISP EDIT ...................................... press 1 time 5.003 DISP F2 .............................. (ADJX,Y) press 2 times 5.004 DISP Adjust CONTROL KNOB for a UUT display of Y= 2047. 5.005 IEEE [@1281] RDG?[I] 5.006 MATH M[1] = MEM 5.007 DISP Adjust CONTROL KNOB for a UUT display of Y=-2047. 5.008 IEEE [@1281] RDG?[I] 5.009 MATH MEM = MEM * -1 5.010 MATH MEM = MEM + M[1] 5.011 MEM* 1000 5.012 ACC 2 500.00mV 12P% 0.2P/ 5.013 MEME 5.014 MEMC mVp-p 9.5U 250mVP 6.001 DISP Set the UUT MENU as follows: 6.001 DISP PARAMATER ................................ press 2 times 6.001 DISP F1 ..................................(AMPL) press 1 time 6.002 DISP Rotate the UUT CONTROL KNOB clockwise until display 6.002 DISP reads "AMP 500.0 mVp". 6.003 DISP Set the UUT MENU as follows: 6.003 DISP EDIT ...................................... press 1 time 6.003 DISP F2 .............................. (ADJX,Y) press 2 times 6.004 DISP Adjust CONTROL KNOB for a UUT display of Y= 2047. 6.005 IEEE [@1281] RDG?[I] 6.006 MATH M[1] = MEM 6.007 DISP Adjust CONTROL KNOB for a UUT display of Y=-2047. 6.008 IEEE [@1281] DCV AUTO,FAST_ON 6.009 IEEE [@1281] RDG?[I] 6.010 MATH MEM = MEM * -1 6.011 MATH MEM = MEM + M[1] 6.012 MEM* 1000 6.013 ACC 2 1000.00mV 12P% 0.2P/ 6.014 MEME 6.015 MEMC mVp-p 23U 500mVP 7.001 DISP Set the UUT MENU as follows: 7.001 DISP PARAMATER ................................ press 2 times 7.001 DISP F1 ..................................(AMPL) press 1 time 7.002 DISP Rotate the UUT CONTROL KNOB clockwise until display 7.002 DISP reads "AMP 2.500 Vp". 7.003 DISP Set the UUT MENU as follows: 7.003 DISP EDIT ...................................... press 1 time 7.003 DISP F2 .............................. (ADJX,Y) press 2 times 7.004 DISP Adjust CONTROL KNOB for a UUT display of Y= 2047. 7.005 IEEE [@1281] RDG?[I] 7.006 MATH M[1] = MEM 7.007 DISP Adjust CONTROL KNOB for a UUT display of Y=-2047. 7.008 IEEE [@1281] RDG?[I] 7.009 MATH MEM = MEM * -1 7.010 MATH MEM = MEM + M[1] 7.011 ACC 20 5.000V 12P% 0.1P/ 7.012 MEME 7.013 MEMC Vp-p 0.095U 2.5VP 8.001 DISP Set the UUT MENU as follows: 8.001 DISP PARAMATER ................................ press 2 times 8.001 DISP F1 ..................................(AMPL) press 1 time 8.002 DISP Rotate the UUT CONTROL KNOB clockwise until display 8.002 DISP reads "AMP 5.000 Vp". 8.003 DISP Set the UUT MENU as follows: 8.003 DISP EDIT ...................................... press 1 time 8.003 DISP F2 .............................. (ADJX,Y) press 2 times 8.004 DISP Adjust CONTROL KNOB for a UUT display of Y= 2047. 8.005 IEEE [@1281] RDG?[I] 8.006 MATH M[1] = MEM 8.007 DISP Adjust CONTROL KNOB for a UUT display of Y=-2047. 8.008 IEEE [@1281] RDG?[I] 8.009 MATH MEM = MEM * -1 8.010 MATH MEM = MEM + M[1] 8.011 ACC 20 10.000V 12P% 0.1P/ 8.012 MEME 8.013 MEMC Vp-p 0.185U 5VP 9.001 HEAD {} 9.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 9.003 HEAD {º Offset Test º} 9.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍͼ} 9.005 HEAD {}ÍÍ Offset Test ÍÍ 9.006 JMP 10.001 9.007 EVAL Dummy 10.001 DISP Set the UUT MENU as follows: 10.001 DISP SETUP .................................... press 3 times 10.001 DISP F2 ................................. (RSET) press 1 time 10.001 DISP F3 ................................. (EXEC) press 1 time 10.002 DISP View the UUT display and verify display reads: 10.002 DISP RESET (Ver 2.2S)X #10.003 DISP Set the UUT MENU as follows: #10.003 DISP EDIT ..................................... press 2 times #10.003 DISP F2 ................................ (ERASE) press 1 time #10.003 DISP F3 ................................. (EXEC) press 1 time #10.004 IEEE [@1281] RDG?[I] #10.005 ACC 1 0.000V 6P% 0.2P/ #10.006 MEME #10.007 MEMC V 0.005U Erase 10.003 DISP Set the UUT MENU as follows: 10.003 DISP PARAMATER ................................ press 2 times 10.003 DISP F1 ................................. (AMPL) press 1 time 10.003 DISP Rotate the UUT CONTROL KNOB clockwise until display 10.003 DISP reads "AMP 10.00 mVp". 10.004 DISP Set the UUT MENU as follows: 10.004 DISP F2 ................................(OFFSET) press 1 time 10.005 DISP Adjust CONTROL KNOB for a UUT display of 5.00V. 10.006 IEEE [@1281] RDG?[I] 10.007 ACC 20 5.000V 6P% 0.1P/ 10.008 MEME 10.009 MEMC V 0.115U 11.001 DISP Adjust CONTROL KNOB for a display of 500 mV. 11.002 IEEE [@1281] RDG?[I] 11.003 ACC 2 500mV 6P% 0.2P/ 11.004 MEM* 1000 11.005 MEME 11.006 MEMC mV 25U 12.001 DISP Adjust CONTROL KNOB for a display of 45.0 mV. 12.002 IEEE [@1281] RDG?[I] 12.003 ACC 200 45.0mV 7P% 0.5P/ 12.004 MEM* 1000 12.005 MEME 12.006 MEMC mV 2.4U 13.001 DISP Adjust CONTROL KNOB for a display of -45.0 mV. 13.002 IEEE [@1281] RDG?[I] 13.003 ACC 200 -45.0mV 7P% 0.5P/ 13.004 MEM* 1000 13.005 MEME 13.006 MEMC mV 2.4U 14.001 DISP Adjust CONTROL KNOB for a display of -500 mV. 14.002 IEEE [@1281] RDG?[I] 14.003 ACC 2 -500mV 6P% 0.2P/ 14.004 MEM* 1000 14.005 MEME 14.006 MEMC mV 25U 15.001 DISP Adjust CONTROL KNOB for a display of -5.00 V. 15.002 IEEE [@1281] RDG?[I] 15.003 ACC 20 -5.000V 6P% 0.1P/ 15.004 MEME 15.005 MEMC V 0.115U 16.001 HEAD 16.002 DISP [27][91]1mDisconnect the test set-up. 16.003 END